Sterilization validation according to ISO 17665-1, EN 13060 and EN 285

Do you need to demonstrate that your sterilization processes – for example, with small or large sterilizers – are repeatable, effective, and can be validated according to standards? HygCen Germany neutrally verifies whether your programs meet the requirements of ISO 17665-1, EN 13060, and EN 285 and whether sterility is guaranteed under real-world conditions – supplemented by international procedures such as ANSI/AAMI ST79, ISO 14937, and ISO 22441.

The validation of sterilization processes is a key component of any standards-compliant reprocessing. ISO 17665-1 describes how steam sterilization procedures must be developed, qualified, and monitored – including test parameters, process design, and performance qualification. For small sterilizers, the device-specific requirements of EN 13060 also apply, and for large sterilizers, EN 285. International standards such as ANSI/AAMI ST79 and ISO 14937 or ISO 22441 supplement the verification framework, particularly for markets outside Europe.

HygCen Germany conducts the validation as a neutral, DAkkS-accredited testing service provider. We do not develop sterilization programs, but exclusively verify whether the processes used by the operator achieve the required performance values ​​and remain reproducible under repeated conditions. This includes process development, OQ/PQ, loading studies, and – if requested by the operator – regular re-validation. Our results are documented in QES-signed test reports and are directly audit- and MDR-compliant.

  • Digital test report – QES-signed
  • Validated reprocessing processes
  • Safe medical devices for users and patients
Request sterilization validation HygCen guarantees you to treat your data confidentially and to use it exclusively in the course of your request.
Herstellerangaben für Medizinprodukte nach en iso 17664

What does validation according to ISO 17665-1 mean in practice?

Validation describes whether a sterilization process:

EN ISO 17665-1 requires a tiered approach for this:

  1. Development of the sterilization process
    → Definition of parameters such as temperature, pressure, steam penetration.
  2. Operational Qualification (OQ)
    → Testing to ensure the device correctly controls the intended parameters.
  3. Performance qualification (PQ)
    → Proof of effectiveness with real or representative loads.
  4. Routine monitoring & re-validation
    → Ensuring long-term process stability.

The different types of devices must always be taken into account:

Loading studies form the core of this validation: HygCen Germany tests whether the loading arrangement is sufficiently heated and penetrated in all areas. This applies particularly to complex instruments, reusable systems, implant-related products, and mixed loads.

Standard criteria – at a glance

Standard criterionRequirements / ContentNotice
Temperature profileDefined temperature-time combinations for steam processesThermosensors, recording, evaluation
Vapor penetrationcomplete penetration in loading areasLoading studies, worst-case configurations
Loading schemesreproducible validation of specific loadsPQ test programs, validated test specimens
Sterilization processISO 17665-1 / 13060 / 285, optional ST79 / 14937 / 22441Process qualification, OQ/PQ
Integrity of the barrierPackaging dependent according to ISO 11607-2 / EN 868 6Barrier and integrity tests
Microbiological requirementsISO 11737-1/-2Bioburden & Sterility
documentationQES-signed resultsAudit-ready report structure

This is how we validate your sterilization processes - step by step

  1. Analysis of the sterilization methods used
    We examine:
     
    • Device type (small sterilizer, large sterilizer, alternative)
    • Programs, media, cycle parameters
    • Loading types (instruments, sets, proximity to implants)
    • existing IFU specifications (ISO 17664)
  2. Test planning & sensor concept
    We define:
     
    • relevant standards (ISO 17665‑1, EN 13060, EN 285, AAMI ST79, ISO 14937 / ISO 22441)
    • Measurement points, measurement methods, repetitions
    • Worst-case loads
    • Validation scenarios
    • Temperature/pressure profiling
  3. Performance Qualification (PQ)
    We assess:
     
    • Steam penetration at critical points
    • Holding times & plateau temperatures
    • Resilience of the programs
    • Loading options & maximum load
    • Compliance with normative requirements
  4. Routine monitoring & re-validation
    We perform:
     
    • Repeat tests after changes
    • annual or event-related re-validations
    • Audit-ready documentation of deviations
    • Temperature profile and sterile goods release check
  5. QES-signed audit report.
    Your report contains:
     
    • Complete process documentation
    • Measured values ​​& curve progressions
    • Evaluation matrix of standard requirements
    • clear statement “Valiable / Not validatable”
    • Interface classification with RDG, cleanliness and packaging

Why HygCen?

FAQ - Frequently Asked Questions

ISO 17665-1 is the overarching standard for steam sterilization processes. It defines the process parameters, validation steps, and verifications required for each type of steam sterilizer. EN 13060 supplements these requirements for small sterilizers, commonly used in medical practices, outpatient clinics, and specialized departments. EN 285, on the other hand, addresses large sterilizers, such as those typically found in central sterile supply departments (CSSDs). The validation principles are identical but differ with regard to load volume, steam generation, cycle design, and sensor technology. HygCen Germany takes these differences into account but verifies only the technical effectiveness - not the process design.

ANSI/AAMI ST79 is the authoritative US standard for steam sterilization and sterile supply. Manufacturers exporting products to the US market, or operators requiring US compliance, benefit from having their sterilization processes additionally tested against these requirements. ST79 specifies expectations regarding steam penetration, loading, sensors, monitoring, and documentation. HygCen Germany can demonstrate whether the sterilization programs used meet not only European but also the relevant US performance criteria. However, this demonstration does not replace regulatory consulting it merely demonstrates the technical performance of the processes.

Steam penetration is one of the crucial effectiveness factors. We test whether the steam can penetrate even hard-to-reach areas, such as hollow bodies, edges, nested loads, or compacted trays. To this end, temperature sensors are placed in critical areas, and worst-case loads are defined. The measurement results must meet the required temperature-time specifications at all measuring points. If deviations are detected, we document them neutrally - adjusting the load or process parameters is the operator's responsibility.

If a cycle fails to meet the normative requirements - for example, insufficient holding times, inadequate steam penetration, or temperature deviations - this is documented neutrally and completely. Common causes can include unfavorable loading configurations, technical malfunctions, unsuitable programs, or altered environmental conditions. HygCen Germany evaluates only the test data; it does not provide process consulting or implement optimizations. The operator decides whether programs need to be adjusted, equipment serviced, or loading modified before a retest is conducted.

Re-validations are required at regular intervals - the exact frequency depends on the quality management system, manufacturer specifications, and operational requirements. Typically, re-validations are performed annually or on an ad-hoc basis, e.g., after repairs, equipment replacements, program changes, new loading concepts, or anomalies detected during monitoring. HygCen Germany fully documents the tests performed but does not specify its own intervals or operational procedures in order to maintain its neutral laboratory role.

Effective sterilization requires that cleaning (ISO 15883), cleanliness (ISO 19227, TOC/Protein/Hemoglobin), and packaging (ISO 11607-2) are already functioning correctly. Sterilization validation is therefore an integral part of the entire reprocessing chain. If upstream processes do not operate reproducibly, sterilization cannot achieve the required level of safety. For this reason, HygCen Germany considers washer-disinfector validation, cleanliness tests, and packaging inspections within a holistic context – without, however, providing consulting services or process design.

REQUEST
PHONE
FIND US

Send request.

* Required fields

Contact us now!

You can reach us under the following phone number:

+49 (0) 385 477 419 00

Call!

You are also welcome to visit us in Schwerin!

You can find us here:

HygCen Germany GmbH
Bornhövedstrasse 78
19055 Schwerin
Germany

Directions!